User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Amprobe RS-3A:Current Clamp DATE: 09-Nov-97 AUTHOR: User Contributed REVISION: 1 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 56 NUMBER OF LINES: 255 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A STANDARD: DTB DJK-50T ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ X 1.003 MMFC C3 1.004 HEAD * INITIAL CONDITIONS * 1.005 DISP The following equipment is need for this procedure: 1.005 DISP Toroid Coil DTB DJK DTB # 5-34 1.006 HEAD {* CURRENT TEST 6 AMP RANGE *} 1.007 JMP 2.001 1.008 EVAL 2.001 DISP Set the UUT FUNCTION/RANGE switch to 6 AMP. 2.002 DISP Reading lock switch (next to function switch) should be 2.002 DISP in the left position. 2.003 DISP Connect a short cabel between the 5725A CURRENT OUTPUT 2.003 DISP HI and LO. 2.004 DISP Clamp the UUT around the center of the loop. 2.005 DISP Adjust the UUT zero control (red adjustment on handel) 2.005 DISP for a UUT zero meter reading. 2.006 5700 6 6.00A 5.00/ 60H B1 2W #! Test Tol 0.3, Sys Tol 0.00257, TUR 116.732 (>= 4.00). 3.001 5700 6 5.00A 5.00/ 60H B1 2W #! Test Tol 0.3, Sys Tol 0.00217, TUR 138.249 (>= 4.00). 4.001 5700 6 4.00A 5.00/ 60H B1 2W #! Test Tol 0.3, Sys Tol 0.00177, TUR 169.492 (>= 4.00). 5.001 5700 6 3.00A 5.00/ 60H B1 2W #! Test Tol 0.3, Sys Tol 0.00137, TUR 218.978 (>= 4.00). 6.001 5700 6 2.00A 5.00/ 60H 2W #! Test Tol 0.3, Sys Tol 0.00134, TUR 223.881 (>= 4.00). 7.001 5700 6 1.00A 5.00/ 60H 2W #! Test Tol 0.3, Sys Tol 0.00069, TUR 434.783 (>= 4.00). 8.001 MESS 8.002 DISP Disconnect UUT from set-up. 8.003 HEAD {* CURRENT TEST 15 AMP RANGE *} 8.004 STD DTB DJK-50T 8.005 JMP 9.001 8.006 EVAL 9.001 DISP [32] Connect the 5725A and the DJK01 as follows: 9.001 DISP [32] [27][91]1m 5725A TO DJK COIL 9.001 DISP [32] CURRENT OUTPUT HI ......... + terminal 9.001 DISP [32] CURRENT OUTPUT LO ......... - terminal 9.002 DISP Set the UUT FUNCTION/RANGE switch to 15 AMP. 9.003 DISP Clamp the UUT around the center of the coil. 9.004 MESS Adjust Stimulus for a UUT reading of 15 Amps. 9.005 5700 0.300A 60H N 2W 9.006 MEM* 50 9.007 MEMC 15 15.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 0.000235, TUR 3191.489 (>= 4.00). 10.001 MESS Adjust Stimulus for a UUT reading of 12 Amps. 10.002 5700 0.240A 60H N 2W 10.003 MEM* 50 10.004 MEMC 15 12.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 0.000196, TUR 3826.531 (>= 4.00). 11.001 MESS Adjust Stimulus for a UUT reading of 9 Amps. 11.002 5700 0.180A 60H N 2W 11.003 MEM* 50 11.004 MEMC 15 9.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 3.1e-005, TUR 24193.548 (>= 4.00). 12.001 MESS Adjust Stimulus for a UUT reading of 6 Amps. 12.002 5700 0.120A 60H N 2W 12.003 MEM* 50 12.004 MEMC 15 6.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 2.2e-005, TUR 34090.909 (>= 4.00). 13.001 MESS Adjust Stimulus for a UUT reading of 3 Amps. 13.002 5700 0.06A 60H N 2W 13.003 MEM* 50 13.004 MEMC 15 3.00A 5.00/ 60H #! Test Tol 0.75, Sys Tol 1.3e-005, TUR 57692.308 (>= 4.00). 14.001 HEAD {* CURRENT TEST 40 AMP RANGE *} 14.002 JMP 15.001 14.003 EVAL 15.001 DISP Set the UUT FUNCTION/RANGE switch to 40 AMP. 15.002 DISP Adjust the UUT zero control (red adjustment on handel) 15.002 DISP for a UUT zero meter reading. 15.003 MESS Adjust Stimulus for a UUT reading of 40 Amps. 15.004 5700 0.800A 60H N 2W 15.005 MEM* 50 15.006 MEMC 40 40.00A 5.00/ 60H #! Test Tol 2, Sys Tol 0.00056, TUR 3571.429 (>= 4.00). 16.001 MESS Adjust Stimulus for a UUT reading of 30 Amps. 16.002 5700 0.600A 60H N 2W 16.003 MEM* 50 16.004 MEMC 40 30.00A 5.00/ 60H #! Test Tol 2, Sys Tol 0.00043, TUR 4651.163 (>= 4.00). 17.001 MESS Adjust Stimulus for a UUT reading of 20 Amps. 17.002 5700 0.400A 60H N 2W 17.003 MEM* 50 17.004 MEMC 40 20.00A 5.00/ 60H #! Test Tol 2, Sys Tol 0.0003, TUR 6666.667 (>= 4.00). 18.001 MESS Adjust Stimulus for a UUT reading of 10 Amps. 18.002 5700 0.200A 60H N 2W 18.003 MEM* 50 18.004 MEMC 40 10.00A 5.00/ 60H #! Test Tol 2, Sys Tol 3.4e-005, TUR 58823.529 (>= 4.00). 19.001 HEAD {* CURRENT TEST 100 AMP RANGE *} 19.002 JMP 20.001 19.003 EVAL 20.001 DISP Set the UUT FUNCTION/RANGE switch to 100 AMP. 20.002 DISP Adjust the UUT zero control (red adjustment on handel) 20.002 DISP for a UUT zero meter reading. 20.003 MESS Adjust Stimulus for a UUT reading of 100 Amps. 20.004 5700 2A 60H N 2W 20.005 MEM* 50 20.006 MEMC 100 100.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.00134, TUR 3731.343 (>= 4.00). 21.001 MESS Adjust Stimulus for a UUT reading of 80 Amps. 21.002 5700 1.6A 60H N 2W 21.003 MEM* 50 21.004 MEMC 100 80.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.00108, TUR 4629.630 (>= 4.00). 22.001 MESS Adjust Stimulus for a UUT reading of 60 Amps. 22.002 5700 1.2A 60H N 2W 22.003 MEM* 50 22.004 MEMC 100 60.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.00082, TUR 6097.561 (>= 4.00). 23.001 MESS Adjust Stimulus for a UUT reading of 40 Amps. 23.002 5700 0.8A 60H N 2W 23.003 MEM* 50 23.004 MEMC 100 40.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.00056, TUR 8928.571 (>= 4.00). 24.001 MESS Adjust Stimulus for a UUT reading of 20 Amps. 24.002 5700 0.4A 60H N 2W 24.003 MEM* 50 24.004 MEMC 100 20.0A 5.00/ 60H #! Test Tol 5, Sys Tol 0.0003, TUR 16666.667 (>= 4.00). 25.001 HEAD {* CURRENT TEST 300 AMP RANGE *} 25.002 JMP 26.001 25.003 EVAL 26.001 DISP Set the UUT FUNCTION/RANGE switch to 300 AMP. 26.002 DISP Adjust the UUT zero control (red adjustment on handel) 26.002 DISP for a UUT zero meter reading. 26.003 MESS Adjust Stimulus for a UUT reading of 300 Amps. 26.004 5700 6A 60H B1 N 2W 26.005 MEM* 50 26.006 MEMC 300 300.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00257, TUR 5836.576 (>= 4.00). 27.001 MESS Adjust Stimulus for a UUT reading of 250 Amps. 27.002 5700 5A 60H B1 N 2W 27.003 MEM* 50 27.004 MEMC 300 250.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00217, TUR 6912.442 (>= 4.00). 28.001 MESS Adjust Stimulus for a UUT reading of 200 Amps. 28.002 5700 4A 60H B1 N 2W 28.003 MEM* 50 28.004 MEMC 300 200.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00177, TUR 8474.576 (>= 4.00). 29.001 MESS Adjust Stimulus for a UUT reading of 150 Amps. 29.002 5700 3A 60H B1 N 2W 29.003 MEM* 50 29.004 MEMC 300 150.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00137, TUR 10948.905 (>= 4.00). 30.001 MESS Adjust Stimulus for a UUT reading of 100 Amps. 30.002 5700 2A 60H B1 N 2W 30.003 MEM* 50 30.004 MEMC 300 100.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00097, TUR 15463.918 (>= 4.00). 31.001 MESS Adjust Stimulus for a UUT reading of 50 Amps. 31.002 5700 1A 60H B1 N 2W 31.003 MEM* 50 31.004 MEMC 300 50.0A 5.00/ 60H #! Test Tol 15, Sys Tol 0.00057, TUR 26315.789 (>= 4.00). 32.001 MESS 32.002 DISP Disconnect UUT from set-up. 32.003 HEAD {* VOLTAGE TEST 150 V RANGE *} 32.004 JMP 33.001 32.005 EVAL 33.001 DISP [32] Connect the 5700A and the UUT as follows: 33.001 DISP [32] [27][91]1m 5700A TO RS-3A 33.001 DISP [32] OUTPUT V Ohm A HI ......... INPUT 33.001 DISP [32] OUTPUT V Ohm A LO ......... INPUT 33.002 DISP Set the UUT FUNCTION/RANGE switch to 150 Volts. 33.003 DISP Adjust the UUT zero control (red adjustment on handel) 33.003 DISP for a UUT zero meter reading. 33.004 5700 150 150.00V 3.00/ 60H 2W #! Test Tol 4.5, Sys Tol 0.013, TUR 346.154 (>= 4.00). 34.001 5700 150 125.00V 3.00/ 60H 2W #! Test Tol 4.5, Sys Tol 0.011, TUR 409.091 (>= 4.00). 35.001 5700 150 100.00V 3.00/ 60H 2W #! Test Tol 4.5, Sys Tol 0.009, TUR 500.000 (>= 4.00). 36.001 5700 150 75.00V 3.00/ 60H 2W #! Test Tol 4.5, Sys Tol 0.007, TUR 642.857 (>= 4.00). 37.001 5700 150 50.00V 3.00/ 60H 2W #! Test Tol 4.5, Sys Tol 0.005, TUR 900.000 (>= 4.00). 38.001 5700 150 25.00V 3.00/ 60H 2W #! Test Tol 4.5, Sys Tol 0.003, TUR 1500.000 (>= 4.00). 39.001 HEAD {* VOLTAGE TEST 300 V RANGE *} 39.002 JMP 40.001 39.003 EVAL 40.001 DISP Set the UUT FUNCTION/RANGE switch to 300 Volts. 40.002 DISP Adjust the UUT zero control (red adjustment on handel) 40.002 DISP for a UUT zero meter reading. 40.003 5700 300 300.00V 3.00/ 60H 2W #! Test Tol 9, Sys Tol 0.028, TUR 321.429 (>= 4.00). 41.001 5700 300 250.00V 3.00/ 60H 2W #! Test Tol 9, Sys Tol 0.024, TUR 375.000 (>= 4.00). 42.001 5700 300 200.00V 3.00/ 60H 2W #! Test Tol 9, Sys Tol 0.017, TUR 529.412 (>= 4.00). 43.001 5700 300 150.00V 3.00/ 60H 2W #! Test Tol 9, Sys Tol 0.013, TUR 692.308 (>= 4.00). 44.001 5700 300 100.00V 3.00/ 60H 2W #! Test Tol 9, Sys Tol 0.009, TUR 1000.000 (>= 4.00). 45.001 5700 300 50.00V 3.00/ 60H 2W #! Test Tol 9, Sys Tol 0.005, TUR 1800.000 (>= 4.00). 46.001 HEAD {* VOLTAGE TEST 600 V RANGE *} 46.002 JMP 47.001 46.003 EVAL 47.001 DISP Set the UUT FUNCTION/RANGE switch to 600 Volts. 47.002 DISP Adjust the UUT zero control (red adjustment on handel) 47.002 DISP for a UUT zero meter reading. 47.003 5700 600 600.00V 3.00/ 60H 2W #! Test Tol 18, Sys Tol 0.052, TUR 346.154 (>= 4.00). 48.001 5700 600 500.00V 3.00/ 60H 2W #! Test Tol 18, Sys Tol 0.044, TUR 409.091 (>= 4.00). 49.001 5700 600 400.00V 3.00/ 60H 2W #! Test Tol 18, Sys Tol 0.036, TUR 500.000 (>= 4.00). 50.001 5700 600 300.00V 3.00/ 60H 2W #! Test Tol 18, Sys Tol 0.028, TUR 642.857 (>= 4.00). 51.001 5700 600 200.00V 3.00/ 60H 2W #! Test Tol 18, Sys Tol 0.017, TUR 1058.824 (>= 4.00). 52.001 5700 600 100.00V 3.00/ 60H 2W #! Test Tol 18, Sys Tol 0.009, TUR 2000.000 (>= 4.00). 53.001 HEAD {* OHMS TEST *} 53.002 JMP 54.001 53.003 EVAL 54.001 DISP Connect the UUT OHMMETER probe (yellow) to the UUT. 54.002 DISP Short the wires from the UUT together at the 5700A. 54.003 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 54.004 DISP [32] Connect the 5700A and the UUT as follows: 54.004 DISP [32] [27][91]1m 5700A TO RS-3A 54.004 DISP [32] OUTPUT V Ohm A HI ......... Ohm INPUT 54.004 DISP [32] OUTPUT V Ohm A LO ......... COM 54.005 ASK+ R K 54.006 ASK- W 54.007 JMP 55.001 54.008 EVAL 55.001 5700 10.00Z +12.5% -11% 2W #! T.U.R. not calculated because tolerance or floor unspecified. 56.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 1 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.